Power quality measurement standard compliance IEEE 1159, IEC 61000-4-30 Ed. 2 Class A
Maximum recording period 1 month (with repeat function on: 55 weeks)
Maximum number of recordable events 1,000 (with repeat function on: 55,000)
Interfaces SD card slot (for writing data to an SD memory card), LAN (for remote operation and downloading of recorded data), USB (for downloading of recorded data), RS-232C (for printer and GPS connectivity)
Display 6.5-inch TFT color LCD (640×480 dots)
Power supply AC ADAPTER Z1002 (12V DC, Rated power supply 100V AC to 24 V AC, 50/60Hz)BATTERY PACK Z1003 (Ni-MH 7.2V DC 4500mAh)
Dimensions, mass 300mm(11.81in)W × 211mm(8.31in)H × 68mm(2.68in)D (excluding protrusions), 2.6kg(91.7oz) (including battery pack)
Supplied Accessories Instruction manual ×1, Measurement guide ×1, VOLTAGE CORD L1000×1 set, Spiral Tube×1, Input Cable Labels ×1, AC ADAPTER Z1002 ×1, Strap ×1, USB cable (1m 3.28ft length) ×1, BATTERY PACK Z1003 ×1, SD MEMORY CARD 2GB Z4001 ×1 Measurement Specifications 1. Time Plot
Voltage items RMS voltage, waveform peak, harmonic, harmonic phase angle, inter-harmonics, high-order harmonic component, total harmonic distortion factor, unbalance factor (negative-phase and zero-phase), DC voltage, frequency Δ V10
Current items RMS current, waveform peak, harmonic, harmonic phase angle, inter-harmonics, high-order harmonic component, total harmonic distortion factor, unbalance factor (negative-phase and zero-phase), DC current, K factor
Power items Active power, reactive power, apparent power, harmonic power, harmonic voltage-current phase difference, active integral power, reactive integral power, power factor 2. Event acquisition Transients, swells, dips, interruptions, frequency, rush current, voltage waveform comparison, timer, external, and other combinations of the time series item events listed above. Input Specifications
Measurement line 1-phase 2-wire, 1-phase 3-wire, 3-phase 3-wire, 3-phase 4-wire
Plus 1 channel in addition to the above measurement lines
Measurement frequency Basic frequency: 50/60/400 HzTransient overvoltage: 2 MHz sampling
Harmonic measurement Orders 0 (DC) to 50 (2.5 kHz or 3 kHz)
Voltage measurement range Voltage measurement: 600.00 Vrm
Transient overvoltage measurement: ±6.0000 kV peak
Current measurement range Load current measurementWith 9694 sensor: 5.0000/50.000 Arms
With 9660 sensor: 50.000/100.00 Arms
With 9661 sensor: 50.000/500.00 Arms
With 9667 sensor: 50.000/500.00 A, 500.00 A/5.0000 kArms
With 9669 sensor: 100.00 A/1.0000 kArms
Leak current measurement With 9675/9657-10 sensor: 500.00 mA/5.0000 Arms
Measurement accuracy (50 Hz/60 Hz measurement) RMS voltage ±0.1% rdg.
RMS current ±0.2% rdg. ±0.1% f.s. + current sensor accuracy
Active power ±0.2% rdg. ±0.1% f.s. + current sensor accuracy
Hãng sản xuất : Hioki - Nhật
Xuất xứ : Nhật