Measurement Accuracy
|
Voltage DC
|
30.000 V
|
0.02%+ 2 counts (upper display)
|
10.000 V
|
0.02%+ 2 counts (lower display)
|
90.00 mV
|
0.02%+ 2 counts
|
-10.00 mV to 75.00 mV
|
0.025 % + 1 count (via TC connector)
|
Current DC
|
24.000 mA:
|
0.02%+ 2 counts
|
Resistance
|
0.0 to 400.0 Ω
|
0.1Ω (4-wire), 0.15Ω (2- and 3-wire)
|
401 to 1500 Ω
|
0.5Ω (4-wire), 1Ω (2- and 3-wire)
|
500 to 3200 Ω
|
1Ω (4-wire), 1.5Ω (2- and 3-wire)
|
Frequency
|
2.0 to 1000.0 CPM
|
0.05 % + 1 count
|
1.0 to 1100.0 Hz
|
0.05 % + 1 count
|
1.00 to 10.00 kHz
|
0.05 % + 1 count
|
Sensitivity
|
1 V peak-to-peak-minimum
|
Pressure
|
Accuracy
|
from 0.025% of range using any of 8 intrinsic safe pressure modules. (for detailed specifications refer to pressure modules in options and accessories)
Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
|